5.7.2.5 Disturbances
When an analogue-active load cell is subjected to the disturbances specified in R 60-2, 2.10.7.5 to 2.10.7.10 (also summarised in Table 5), the difference between the load cell output due to a disturbance and the load cell output without disturbance (fault) shall satisfy the conditions in 5.7.1.1.
Test | Subclause R 60-2, 2.10 test procedure | Characteristic under test | |
|---|---|---|---|
Warm-up time | 2.10.7.3 | 1.0 | Influence factor |
Power voltage variations | 2.10.7.4 | Influence factor | |
Short-time power reductions | 2.10.7.5 | Disturbance | |
Bursts (electrical fast transients) | 2.10.7.6 | Disturbance | |
Surge | 2.10.7.7 | Disturbance | |
Electrostatic discharge | 2.10.7.8 | Disturbance | |
Electromagnetic susceptibility | 2.10.7.9 | Disturbance | |
Immunity to conducted electromagnetic fields | 2.10.7.10 | Disturbance | |
Span stability | 2.10.7.11 | Influence factor |