5.7.2.5 Disturbances

When an analogue-active load cell is subjected to the disturbances specified in R 60-2, 2.10.7.5 to 2.10.7.10 (also summarised in Table 5), the difference between the load cell output due to a disturbance and the load cell output without disturbance (fault) shall satisfy the conditions in 5.7.1.1.

Table 5 — Influence factors and disturbances

Test

Subclause R 60-2, 2.10 test procedure

a)

Characteristic under test

Warm-up time

2.10.7.3

1.0

Influence factor

Power voltage variations

2.10.7.4

Influence factor

Short-time power reductions

2.10.7.5

Disturbance

Bursts (electrical fast transients)

2.10.7.6

Disturbance

Surge

2.10.7.7

Disturbance

Electrostatic discharge

2.10.7.8

Disturbance

Electromagnetic susceptibility

2.10.7.9

Disturbance

Immunity to conducted electromagnetic fields

2.10.7.10

Disturbance

Span stability

2.10.7.11

Influence factor