8.5 Disturbances

The MPE of the gas analytical system shall be in compliance with 4.3.2. Significant faults shall be detected and fixed by means of checking facilities during the following tests, carried out to verify the requirements of 4.5.4 under rated operating conditions (as specified in 4.5.1):

  1. mechanical shocks: …​…​…​…​…​…​…​…​…​…​…​ R 144-2, 1.11.2;

  2. short time power reduction: …​…​…​…​…​…​…​…​ R 144-2, 1.12;

  3. bursts from the mains (transients): …​…​…​…​…​ R 144-2, 1.13;

  4. electrostatic discharges: …​…​…​…​…​…​…​…​.. R 144-2, 1.14;

  5. radio frequency electromagnetic fields: …​…​…​…​ R 144-2, 1.15.