2.1.4 Temperature effects on minimum dead load output (MDLO) ( Change MDLO)
2.1.4.1
Enter in Table 5 the average indication for the initial minimum test load, , for each test temperature from Table 1.
2.1.4.2
Calculate the difference between the average test indications for each temperature in sequence and divide the result by the conversion factor, , to obtain the change in terms of the load cell verification interval, .
2.1.4.3
Divide by () and multiply the result by a factor for class B, C, and D load cells or for class A load cells. This gives the change in per for class B, C, and D load cells or in per for class A load cells.
2.1.4.4
Multiply the result by ] to give the final result in units of per for class B, C, and D load cells, or in units of per for class A load cells. must not exceed .