Second edition

Date: 2021-10-01

TC 9

Metrological regulation for load cells —
Part 3: Test report format
Réglementation métrologique des cellules de pesée —
Partie 3: Format du rapport d’essais


OIML Publications may be downloaded from the OIML web site in the form of PDF files. Additional information on OIML Publications may be obtained from the Organisation’s headquarters:

Bureau International de Métrologie Légale
11, rue Turgot – 75009 Paris – France
Telephone: 33 (0)1 48 78 12 82
Fax: 33 (0)1 42 82 17 27
E-mail: biml@oiml.org
Internet: www.oiml.org



Foreword

The International Organisation of Legal Metrology (OIML) is a worldwide, intergovernmental organisation whose primary aim is to harmonise the regulations and metrological controls applied by the national metrological services, or related organisations, of its Member States. The main categories of OIML publications are:

  • International Recommendations (OIML R), which are model regulations that establish the metrological characteristics required of certain measuring instruments and which specify methods and equipment for checking their conformity. OIML Member States shall implement these Recommendations to the greatest possible extent;

  • International Documents (OIML D), which are informative in nature and which are intended to harmonise and improve work in the field of legal metrology;

  • International Guides (OIML G), which are also informative in nature and which are intended to give guidelines for the application of certain requirements to legal metrology;

  • International Basic Publications (OIML B), which define the operating rules of the various OIML structures and systems; and

OIML Draft Recommendations, Documents and Guides are developed by Project Groups linked to Technical Committees or Subcommittees which comprise representatives from OIML Member States. Certain international and regional institutions also participate on a consultation basis. Cooperative agreements have been established between the OIML and certain institutions, such as ISO and the IEC, with the objective of avoiding contradictory requirements. Consequently, manufacturers and users of measuring instruments, test laboratories, etc. may simultaneously apply OIML publications and those of other institutions.

International Recommendations, Documents, Guides and Basic Publications are published in English (E) and translated into French (F) and are subject to periodic revision.

Additionally, the OIML publishes or participates in the publication of Vocabularies (OIML V) and periodically commissions legal metrology experts to write Expert Reports (OIML E). Expert Reports are intended to provide information and advice, and are written solely from the viewpoint of their author, without the involvement of a Technical Committee or Subcommittee, nor that of the CIML. Thus, they do not necessarily represent the views of the OIML.

This publication — reference OIML R 60-3:2021 — is an updated edition (developed by the OIML Certification System Management Committee) of R 60-3:2017 (developed by Project Group 1 of OIML Technical Committee TC 9 Instruments for measuring mass and density). This updated edition consolidates the Amendment (2019-12-23) to R 60:2017, and includes other editorial and minor technical changes. It was approved for final publication by the International Committee of Legal Metrology at its 56th meeting in October 2021 and was sanctioned by the 16th International Conference on Legal Metrology in 2021. It supersedes the previous edition of R 60 dated 2017.

OIML Publications may be downloaded from the OIML web site in the form of PDF files. Additional information on OIML Publications may be obtained from the Organisation’s headquarters:

Bureau International de Métrologie Légale
11, rue Turgot — 75009 Paris — France
Telephone: 33 (0)1 48 78 12 82
Fax:33 (0)1 42 82 17 27
E-mail:biml@oiml.org
Internet: www.oiml.org

Metrological regulation for load cells — Part 3:Test report format

1  Introduction

1.1 

This Report Format applies to any kind of load cell (independent of its technology). It presents a standardised format for the results of the various tests and examinations, described in OIML R 60-2, to which a type of load cell shall be submitted for the purpose of its approval based on this OIML Recommendation.

1.2 

It is recommended that all metrology services or laboratories evaluating and/or testing types of load cells according to OIML R 60-1, or to national or regional regulations based on this Recommendation, use this Report Format, directly or after translation into a language other than English or French. In the case of a translation, it is highly recommended to leave the structure and the numbers of the clauses unchanged: in this case, most of the content is also understandable for those who cannot read the language of the translation.

1.3 

Some of the tests may have to be repeated several times and reported using several identical sheets; therefore, report pages must be numbered in the space provided at the top of each page, completed by the indication of the total number of pages.

1.4 

In the practical application of the Report Format, in addition to a cover page by the Issuing Authority, as a minimum, clauses A-F (as necessary) shall be included.

2  Applicability of this Report Format

In the framework of OIML B 18 Framework for the OIML Certification System (OIML-CS) [6] applicable to load cells in conformity with OIML R 60-1 and R 60-2, use of this Report Format is mandatory, in French and/or in English with translation into the national languages of the countries issuing such certificates, if appropriate.

Implementation of this Report Format is informative with regard to the implementation of OIML R 60-1 and R 60-2 in national regulations.

2.1  Calculation procedures

2.1.1 

In order to facilitate a comparison of the reports established in English and in French, the same abbreviations (those of the English language) are used in both versions; the meanings of these abbreviations are given whenever appropriate.

In testing and evaluating load cells for type evaluation, it is recognised that the test apparatus and practices used by the various laboratories will be different. OIML R 60 allows for these variations and still provides a method for testing, recording and calculating results that are readily understandable by other knowledgeable parties reviewing the data.

In order to achieve this ease of comparability it is necessary that those persons conducting the tests use a common system for recording data and calculating results.

Thus, it is essential that the calculation procedures below be reviewed and followed closely in the completion of this test report.

2.1.2  Load cell errors ( E L = Error Load test)

2.1.2.1 

Complete a Table 1 for each test temperature, calculate the averages and record in the right-hand column. When five runs are necessary, use Table 2.

2.1.2.2 

Determine the conversion factor, f , which is the number of indicated units per load cell verification interval, v , and is used to convert all “indicated units” to “ v ”. It is determined from the test data averages of the increasing load tests at the initial 20 °C nominal test temperature.

2.1.2.3 

If a test load corresponding to 75 % of the measuring range for the load cell under test (i.e. 2 250 divisions for a 3 000 division cell, which is D min plus 75 % of the difference between D max and D min ) is not included in the test loads used in Table 1, interpolate between the adjacent upper and lower values of the averages of all three test runs and record in Table 3 (see R 60-2, 2.8.2).

2.1.2.4 

Calculate the difference between the average indication on the increasing load test runs at 75 % of the difference between D max and D min and the indication at D min . Divide the result (to five significant figures) by the number of verification intervals (75 % of n ) for that load to obtain the conversion factor, f , and record in the tables that follow.

f = average indication at    0.75 ( D max D min ) indication at    D min 0.75 n

The units of conversion factor, f , are indicated units (e.g. digits or counts) per load cell verification interval, v .

2.1.2.5 

Enter the average test indications of the tests at the temperatures following the initial test at a nominal 20 °C in Table 3. In recording this data, indicate a “no test load” indication (at D min ) as “0”. This may require subtracting the “no load indication at D min ” from the “test load indication” so that the first entry in the column is “0”. These “0’s” have been preprinted on the form to clarify that a dead load condition is recorded as “0”.

2.1.2.6 

Calculate the reference indication, R i , by converting the net test load, in mass units, to indicated units (e.g. counts or digits), by multiplying by the conversion factor, f , for each test load and recording in the 2nd column in Table 3.

R i = ( test load    i D min ) ( D max D min ) n f

2.1.2.7 

In Table 3 calculate the difference between the average test indication and the reference indication for each test load at each test temperature and divide the result by the conversion factor, f , to obtain the error, E L , for each test load in terms of v .

E L = ( average test indication for test load    i reference indication    R i ) / f

2.1.2.8 

Compare E L with the corresponding MPE for each test load.

2.1.3  Repeatability error ( E R in terms of verification interval, v )

2.1.3.1 

Enter data in Table 4.

2.1.3.2 

Calculate the maximum difference between the test indications on Form 6.3 and divide it by f to obtain the repeatability error, E R , in terms of the load cell verification interval, v .

E R = ( maximum indication of the test load minimum indication ) / f

2.1.3.3 

Compare E R with the absolute value of the corresponding MPE for each test load.

2.1.4  Temperature effects on minimum dead load output (MDLO) ( C M = Change MDLO)

2.1.4.1 

Enter in Table 5 the average indication for the initial minimum test load, D min , for each test temperature from Table 1.

2.1.4.2 

Calculate the difference between the average test indications for each temperature T i in sequence and divide the result by the conversion factor, f , to obtain the change in terms of the load cell verification interval, v .

C M = ( average test indication at    T 2 average indication at    T 1 ) / f

2.1.4.3 

Divide C M by ( T 2 T 1 ) and multiply the result by a factor T f = 5 for class B, C, and D load cells or T f = 2 for class A load cells. This gives the change in v per 5 °C for class B, C, and D load cells or in v per 2 °C for class A load cells.

2.1.4.4 

Multiply the result by [ ( D max D min ) / n / v min ] to give the final result C M ( v min ) in units of v min per 5 °C for class B, C, and D load cells, or in units of v min per 2 °C for class A load cells. C M ( v min ) must not exceed p LC .

C M ( v min ) = | C M T f ( T 2 T 1 ) × ( D max D min ) n v min |

C M ( v min ) p L C

2.1.5  Creep magnitude C C ( t ) and minimum dead load output return ( C DR )

( C C ( t ) = Creep, expressed in terms of the load cell verification interval, v )

( C D R = D R , expressed in terms of the load cell verification interval, v )

Remark: Contrary to the minimum dead load output return D R in terms of mass the minimum dead load output C D R is expressed in terms of the verification interval, v .

From the test indications recorded in Table 6, calculate the difference between the initial indication obtained at the minimum creep test load after the stabilisation period and any indication obtained over the 30 minute test period with the maximum creep test load of 90 % to 100 % of E max and divide by the conversion factor, f .

C C ( t ) = ( indication initial indication ) / f

Remark: If the minimum creep test load or the maximum creep test load differ from D min or D max according to 2.1.2 “Load cell errors E L ” the conversion factor, f , must be recalculated with the minimum and maximum creep test loads (see 2.1.2.4).

2.1.5.1 

C C ( t ) must not exceed 0.7 times the absolute value of the MPE for the maximum creep test load at any time t over the 30 minute creep test period.

2.1.5.2 

Calculate the difference between the test indications obtained at t = 20 minutes and t = 30 minutes after the initial indication at t = t 0 and divide by f to obtain the creep error, C C ( 30 20 ) , in terms of the load cell verification interval, v .

C C ( 30 20 ) = ( indication at time    t = 30    minutes indication at time    t = 20    minutes ) / f

2.1.5.3 

C C ( 30 20 ) shall not exceed 0.15 times the absolute value of the MPE for the applied load.

2.1.5.4 

Calculate the difference between the initial indication obtained at the minimum creep test load after the stabilisation period ( t 0 = 0 min ) and the indication at the minimum creep test load after the creep test and after a time interval for stabilisation ( t > 30 min ) and divide the result by conversion factor, f , to obtain the minimum dead load output return, C DR , in terms of v .

C DR = ( minimum test load indication 2 minimum test load indication 1 ) / f

2.1.5.5 

If the time intervals specified in R 60-2, Table 1 have been met, C DR must not exceed 0.5 v .

2.1.5.6 

If the actual time is between 100 % and 150 % of the specified time in R 60-2, Table 1, then the following applies:

C DR 0.5 ( 1 ( x 1 ) ) v

with

x = actual time/specified time

2.1.5.7 

Whereas C DR expresses the minimum dead load output return in terms of v , the value of D R as used in OIML R 76 [1] is expressed in units of mass ( g , kg or t ).

2.1.5.8 

Calculate the minimum dead load output return, D R , expressed in units of mass ( g , kg or t ) as follows: D R = ( E max E min ) C DR / n LC

2.1.5.9 

Regardless of the value declared by the manufacturer for the apportionment factor, p LC , the MPE for creep shall be determined from R 60-1, Table 4 using the apportionment factor, p LC = 0,7 (see R 60-1, 5.5.1).

2.1.6  Barometric pressure effects 1) ( C P = Change due to barometric pressure)

2.1.6.1 

From the test indications recorded in R 60-3, Table 7, calculate the difference between the indications for each pressure and divide the result by conversion factor, f , to obtain the change, C P , in terms of v .

C P = ( indication at  P 2 indication at  P 1 ) / f

2.1.6.2 

Divide C P by ( P 2 P 1 ) to determine the change due to barometric pressure in terms of v per kilopascal ( kPa ).

2.1.6.3 

Multiply the result by [ ( E max E min ) / n LC to obtain the result in units of mass (g, kg, or t) per kPa (as stated by the manufacturer). The result must not exceed v min .

C P ( v ) = C P ( P 2 P 1 ) ( E max E min ) n max v min

2.1.7  Humidity effects 2) (CH or no mark)

( C Hmin = Change in terms of v due to Humidity effect on the indication of the minimum test load D min )

( C Hmax = Change due to Humidity effect on the indication of the maximum test load D max )

Remark: If the minimum or maximum test load used for this test differ from the minimum test load D min or maximum test load D max according to R 60-3, 2.1.2 “Load cell errors E L ” the conversion factor, f , must be recalculated with the minimum and maximum test loads of this test (see R 60-3, 2.1.2.4).

2.1.7.1 

From the test indications recorded in R 60-3, Table 6.10.1, calculate the difference between the initial indications for the minimum test load, D min , before and after the damp heat test and divide the result by conversion factor, f , to obtain the change, C Hmin , in terms of verification interval, v (see R 60-1, 5.6.3.1).

C Hmin = [ ( indication at  D min ) after ( indication at  D min ) before ] / f

C Hmin  must not exceed 0.04  n .

2.1.7.2 

Calculate the average indications I ¯ { D max } and I ¯ { D min at D min (see R 60-2, 2.10.5) for the required number of test indications, before and after the damp heat test. Subtract I ¯ { D min from I ¯ { D max } for the tests before and after damp heat test and then calculate the difference between the results. Divide the result by the conversion factor, f , to obtain the change, C Hmax in terms of v .

C Hmax = [ ( I ¯ { D max } I ¯ { D min } ) after ( I ¯ { D max } I ¯ { D min } ) before ] f

2.1.7.3 

C Hmax must not exceed the MPE (see R 60-1, Table 4 in 5.3.2).

2.1.8  Humidity effects 3) (SH)

Report load test errors at different temperatures and humidity conditions using R 60-3, Forms 6.3, then indicate the results in R 60-3, Table 6.10.2 utilising the procedure contained within “load cell errors” procedure, R 60-3, 2.1.2, in a manner similar to that used for the preparation of R 60-3, Table 6.5.

2.2  Additional tests for digital load cells

2.2.1  Warm-up time

2.2.1.1 

Enter data on R 60-3, Form 6.11 (Warm-up time).

2.2.1.2 

Span is the result of subtraction of the indication at the minimum test load, D min , from the indication at the maximum test load, D max .

2.2.1.3 

Change is the difference between the span and the initial run span.

2.2.2  Power voltage variations

2.2.2.1 

Enter data on R 60-3, Form 6.12.

2.2.2.2 

Perform load tests and record results utilising R 60-3, Form 6.12.

2.2.2.3 

Calculate reference indications in accordance with the “load cell errors” procedure, R 60-3, 2.1.2.

2.2.2.4 

Indicate results on R 60-3, Form 6.12.

2.2.3  Short-time power reductions

2.2.3.1 

Enter data on R 60-3, Form 6.13.

2.2.3.2 

Calculate the difference, which is:

difference = ( indication with disturbance, in units indication without disturbance, in units ) conversion factor , f

2.2.3.3 

Indicate results on R 60-3, Form 6.13.

2.2.4  Bursts (electrical fast transients)

2.2.4.1 

Enter data on R 60-3, Forms 6.14.1 and 6.14.2.

2.2.4.2 

Indicate results on R 60-3, Forms 6.14.1 and 6.14.2.

2.2.5  Surge

2.2.5.1 

Enter data on R 60-3, Forms 6.15

difference = ( indication with disturbance, in units indication without disturbance, in units ) conversion factor,  f

2.2.5.2 

Indicate results on R 60-3, Forms 6.15

2.2.6  Electrostatic discharge

2.2.6.1 

Enter data on R 60-3, Forms 6.16.1, 6.16.2 and 6.16.3.

2.2.6.2 

Calculate the difference, which is:

difference = ( indication with disturbance, in units indication without disturbance, in units ) conversion factor , f

2.2.6.3 

Indicate results on R 60-3, Forms 6.16.1, 6.16.2.1, and 6.16.2.2.

2.2.6.4 

Provide test point information on Form 6.16.3.

2.2.7  Electromagnetic susceptibility

2.2.7.1 

Enter data on R 60-3, Form 6.17.1.

2.2.7.2 

Calculate the difference, which is:

difference = ( indication with disturbance, in units indication without disturbance, in units ) conversion factor , f

2.2.7.3 

Indicate results on R 60-3, Form 6.17.1.

2.2.7.4 

Provide test set-up information on R 60-3, Form 6.17.2.

2.2.8  Immunity to conducted electromagnetic fields

2.2.8.1 

Enter data on R 60-3, Form 6.18.

2.2.8.2 

Calculate the difference, which is:

difference = ( indication with disturbance, in units indication without disturbance, in units ) conversion factor , f

2.2.8.3 

Indicate results on R 60-3, Form 6.18.

2.2.8.4 

Provide test setup information on R 60-3, Form 6.18.

2.2.9  Span stability

2.2.9.1 

Enter data on R 60-3, Forms 6.19.1 (3 runs) to 6.19.2 (5 runs).

2.2.9.2 

Calculate averages and record on R 60-3, Forms 6.19.1 (3 runs) to 6.19.2 (5 runs).

2.2.9.3 

Indicate results on R 60-3, Form 6.19.3

2.3  General notes

2.3.1 

Absolute (not relative) time shall be recorded.

Note  For example, Figure 1 below gives a sample plot depicting the combined errors versus applied load.

2.3.2 

When reporting values for individual test data, the data should be truncated to two significant digits to the right of the decimal place and reported in load cell verification intervals, v .

Figure 1 — Example of an error envelope

2.4  Formula signs and list of symbols

SymbolDescriptionReference
C C ( t ) creep magnitude, expressed in terms of v at time t obtained over the 30 minutes creep test.2.1.5
C C ( 30 20 ) difference between output at t = 30 minutes and at t = 20 minutes during creep test.2.1.5.2
C DR minimum dead load output return, expressed in terms of v 2.1.5
C Hmax humidity effect on maximum test load output, expressed in terms of v 2.1.7
C Hmin humidity effect on minimum test load output, expressed in terms of v 2.1.7
C M temperature effect on minimum dead load output, expressed in terms of v .2.1.4
C M ( v min ) temperature effect on minimum dead load output, expressed in terms of v min per 5 °C for class B, C and D or per 2 °C for class A.2.1.4
C P barometric pressure effect, expressed in terms of v 2.1.6
C P ( v min ) barometric pressure effect, expressed in terms of mass ( g , kg , t ) per kPa .2.1.6
D max maximum test loadR 60-1, 3.5.6
D min minimum test loadR 60-1, 3.5.12
D R minimum dead load output return, expressed in mass units ( g , kg , t )R 60-1, 3.5.10
E L load cell error, expressed in terms of v 2.1.2
E max maximum capacity of the load cellR 60-1, 3.5.5
E min minimum dead load of the load cellR 60-1, 3.5.9
E R repeatability error, expressed in terms of v 2.1.3
f conversion factor, number of indicated units per verification interval, v 2.1.2.4
MPEmaximum permissible errorR 60-1, 3.7.10
n number of load cell verification intervals into which the load cell measuring range is dividedR 60-1 3.5.13
n LC maximum number of load cell verification intervalsR 60-1, 3.5.8
p LC apportionment factorR 60-1, 3.7.2
R i reference indication (net test load), expressed in indication units2.1.2.6
t 0 time t 0 = 0 min when the initial indication at minimum test load is measured2.1.5
t time over the 30 minute creep test period after the initial indication ( t 0 = 0 min ) at minimum test load2.1.5
T 1 , T 2 temperature1, temperature22.1.4.2
v load cell verification intervalR 60-1, 3.5.4
v min minimum load cell verification intervalR 60-1, 3.5.11
Y relative v min , Y = ( E max E min ) / v min R 60-1, 3.5.15,
Z relative DR, Z = ( E max E min ) / ( 2 × D R ) R 60-1, 3.5.14

2.5  Summary of formulae contained within calculation procedures

SymbolFormula
C C C C = ( indication initial indication ) / f
C C ( 30 20 ) C C ( 30 20 ) = ( test indication at 30 minutes test indication at 20 minutes ) / f
C DR C DR = ( minimum test load indication2 minimum test load indication1 ) / f
C Hmin C Hmin = [ ( indication at    D min ) after ( indication at    D min ) before // f]
C Hmax C Hmax = [ ( indication at    D max indication at    D min ) after ( indication at    D max indication
C M C M = ( average test indication at    T 2 average indication at    T 1 ) / f
C P C P = ( indication at    P 2 indication at    P 1 ) / f
DR D R = ( E max E min ) × C DR / n LC
E L E L = ( average test indication reference indication ) / f
E R E R = ( maximum indication minimum indication ) / f
f f = average indication at  0,75 ( D max D min ) indication at  D min / 0,75 n [see Note 2]
R i R i = [ ( test load D min ) / ( D max D min ) ] × n × f

Note 1  Observe extreme caution by referring to calculation procedure for correct application of these formulae.

Note 2  Use with initial 20 °C ascending load run only. Refer to R 60-2, 2.8.2.

3  Guidance for the application of this Test Report Format

In case a prescribed test is not relevant for the type of instrument to be tested, the reason why the test is omitted shall be clearly stated in the field “Remarks” (for instance surge tests on signal lines shorter than 30 m, tests related to AC mains supply in case of an instrument only powered by batteries, or partial testing after modification of a previously tested type).

The number of the report and the page numbers shall be completed in the heading.

Page 1 of this Report Format may be replaced by a cover page by the Issuing Authority.

Enter “NA” or “/” for “the test is not applicable”.

4  The Evaluation Report

Cover page by the Issuing Authority


4.1  Authority responsible for this Report

Name
Address
Report number
Application number
Period of tests
Date of issuing this Report
Name and signature of the
responsible person
Stamp(s) (if applicable)

4.2  Synopsis of the results of the examination and tests

The load cell under test fulfils ALL the applicable requirements according to OIML R 60-1

Yes

No

Remarks:

4.3  Summary of the results of the examination and tests

EDITORIAL NOTE — To be completed by the Issuing Authority

4.3.1  Examinations

For details, refer to the tests as indicated in the last column.

General requirementsPassedFailedDetails in R 60
DocumentationR 60-2, 2.5
Inscription and presentation of load cell informationR 60-1, 6.2

4.3.2  Performance tests (Refer to R 60-2, 2.10)

For details, refer to the tests as indicated in the last column.

Tests performed at ( 20 °C / X 1 °C / X 2 °C / 20 °C ):

Test procedurePassedFailedDetails in R 60
Maximum permissible measurement errorsR 60-1, 5.3 / R 60-2, 2.10.1
Repeatability errorR 60-1, 5.4 / R 60-2, 2.10.1
Temperature effect on minimum dead load output returnR 60-1, 5.6.1.3 / R 60-2, 2.10.1
Creep testR 60-1, 5.5.1 / R 60-2, 2.10.2
Minimum dead load output return (DR)R 60-1, 5.5.2 / R 60-2, 2.10.3
Barometric pressure effects at ambient temperatureR 60-1, 5.6.2 / R 60-2, 2.10.4
Humidity effects (CH, SH)R 60-1, 5.6.3 / R 60-2, 2.10.5 / 2.10.6

Additional tests performed for digital load cells:

Test procedurePassedFailedDetails in R 60
Warm-up timeR 60-1, 5.7.2.1 / R 60-2, 2.10.7.3
Power voltage variationsR 60-1, 5.7.2.2 / 5.7.2.3 / 5.7.2.4 / R 60-2, 2.10.7.4
Short-time power reductionsR 60-1, 5.7.2.5 / R 60-2, 2.10.7.5
Bursts (electrical fast transients)R 60-1, 5.7.2.5 / R 60-2, 2.10.7.6
SurgeR 60-1, 5.7.2.5 / R 60-2, 2.10.7.7
Electrostatic dischargeR 60-1, 5.7.2.5 / R 60-2, 2.10.7.8
Electromagnetic susceptibilityR 60-1, 5.7.2.5 / R 60-2, 2.10.7.9
Immunity to conducted electromagnetic fieldsR 60-1, 5.7.2.5 / R 60-2, 2.10.7.10
Span stabilityR 60-1, 5.7.2.6 / R 60-2, 2.10.7.11
SoftwareR 60-1, 6.1

4.4  General information regarding the evaluation process

4.4.1  Manufacturer of the specimen

Company
Address
Contact information

4.4.2  Applicant

Company

Representative (name, telephone)

Address

Contact information

Reference

Date of application

Application number

Applicant authorised by the manufacturer (documented)

☐ Yes

☐ No

Statement that no concurrent application for OIML type evaluation has been made to any other OIML Issuing Authority (see Procedural Document PD-05 [29], 4.1.2 b)

☐ Yes

☐ No

Remarks:

4.4.3  Testing laboratories involved in the tests

(This table has to be completed for each test laboratory)

Name

Address

Application number

Tests by this laboratory

Date/period of tests

Name(s) of test engineer(s)

Accredited by

Number:

Expires (date):

Accreditation includes R 60

☐ Yes

Edition: ☐

☐ No

Details of relevant peer assessment or assessment by other means

In case tests have been performed at locations other than the address of this laboratory, give details here

Name of the responsible person

Date of signature

Stamp (if applicable) and signature of the responsible person

Remarks: ‌

4.5  General information concerning the load cell type

(as provided by the manufacturer prior to the evaluation)

Manufacturer’s name/trade mark
Manufacturer’s type designation (or load cell model number)
UnitRange
Accuracy classes
Maximum number of verification intervals, n LC
Maximum capacity, E max ( g , kg , t )
Minimum capacity, E min ( g , kg , t )
Minimum load cell verification interval, v min = ( E max E min ) / Y ( g , kg , t )
Minimum dead load output return, D R = 1 / 2 ( E max E min ) / Z ( g , kg , t )
Rated output( mV V 1 or counts)
Input impedance

4.6  Accessories, supplied with the test pattern by the applicant

AccessoryRemarks and specifications
Analogue data processing device (see OIML R 76 [1], T.2.2.3)
Cables
Load cell mounting hardware
Load introduction elements
Main power supply
Battery (type, voltage)
Indicator (see OIML R 76 [1], T.2.2.2)
Data printer
Other accessories:

Further remarks concerning accessories:

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

4.7  Selection of sample(s) tested

4.7.1  Definition of the test pattern (supplied by the applicant for this test report)

This test report is issued for the following load cell:

Model designationSerial numberMaximum capacityMaximum number of load cell intervalsMinimum load cell verification intervalMinimum dead load output return

E max
( g , kg , t )

n LC

v min
( g , kg , t )

DR ( g , kg , t )

4.7.2  Justification of the selection of the test sample(s)

(refer to R 60-2, 2.3, 2.4 and Annex D):

Model designationSerial numberJustification / RemarkTest Report No. (if available)

4.8  Adjustments and modifications made to the samples during the testing:

Justification of the selection of the test sample(s) (refer to R 60-2, 2.3):

Model designationSerial numberAdjustments and modifications made to the samplesTest Report No.
(if available)

Further information concerning adjustments:

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

4.9  Additional information concerning the type

4.9.1  General information of the load cell under test (specified by the manufacturer)

Manufacturer’s name/trade mark
Manufacturer’s type designation (or load cell model number)
Serial number
Load cell construction (e.g. S-type, ring type, bending beam)
Load cell material
Sealing of strain gauge application (e.g. hermetically, potted)
Digital load cell (Yes / no)
Accuracy classes
Maximum number of verification intervals, n LC
Maximum capacity, E max ( g , kg , t )
Minimum capacity, E min ( g , kg , t )
Minimum load cell verification interval, v min = ( E max E min ) / Y ( g , kg , t )
Minimum dead load output return, DR = 1 2 ( E max E min ) / Z ( g , kg , t )
Rated output( mV V 1 or counts)
Input impedancea
Cable connectiona4-wire / 6-wire
Cable lengthb m

a  mandatory for strain gauge load cells

b  mandatory for strain gauge load cells with 4-wire connection

Additional information concerning the type (connection equipment, interfaces, etc.):

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

4.9.2  Additional information for the performance tests

(Refer to R 60-1, 6.2.2, 6.2.3, and 6.2.4)

Accuracy class ☐ A ☐ B ☐ C ☐ D
Working temperature (if other than 10 °C to + 40 °C ): Upper_____ °C , Lower_____ °C
Humidity symbol ☐ NH ☐ SH ☐ CH or no marking
Loading designation: (refer to R 60-1, 6.2.4.2) Tension Compression Universal Beam (shear) Beam (bending)
Minimum dead load as: E min =
Safe load limit as: E lim =
Excitation voltage: ☐ AC ☐ DC
Value of the apportionment factor, p LC , if not equal to 0.7

4.9.3  Additional information of the test pattern for digital load cells

Power voltage: ☐ AC ☐ DC
Interfaces:
Output signal:
Software identification:
Value of the apportionment factor, p LC , if not equal to 0.7

4.9.5  Documentation supplied with the test pattern by the applicant

Name of the documentContentVersion No. / date of issue

4.9.6  Inscriptions and presentations of load cell information

(according to manufacturer statement, refer to R 60-1, 6.2)

R 60-1 referenceInformationOn the load cellAccompanying documentIn the data sheet
6.2.1 / 6.2.2Name or trademark of manufacturer
6.2.1 / 6.2.2Manufacturer’s own designation or load cell model
6.2.1Serial numberNot applicable
6.2.1Year of productionNot applicable
6.2.2 / 6.2.4.1Accuracy class(es) and their symbols
6.2.4.5Maximum number of load cell verification intervals, n LC
6.2.2 / 6.2.4.2Type of load
6.2.2 / 6.2.4.3Working temperature designation
6.2.2 / 6.2.4.4Humidity symbol “NH”
6.2.2 / 6.2.4.4Humidity symbol “SH”
6.2.2 / 6.2.4.4No humidity symbol or “CH”
6.2.2Minimum dead load, E min a
6.2.1 / 6.2.2Maximum capacity, E max a
6.2.2Safe load limit, E lim a
6.2.2Minimum load cell verification interval ( v min )a
6.2.3, aRelative v min ( Y )
6.2.3, bMinimum dead load output return DRa
6.2.3, bRelative DR ( Z )
6.2.2, lRated output
6.2.2, lExcitation voltage
6.2.2, lInput impedance
6.2.2, lCable connectionb
6.2.2, lCable lengthc
6.2.2, kApportionment factor, p LC (if not equal to 0.7)
6.2.2, l 6.2.3, cFurther information

a  in units of ( g , kg , t )

b  e.g. 4-wire / 6-wire cable

c  mandatory for strain gauge load cells with 4-wire connection

Further load cell information given by the manufacturer:

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

4.9.7  Various designs within the model range:

Model designationMaximum capacityMinimum dead loadMaximum number of load cell intervalsMinimum load cell verification intervalMinimum dead load output return
E max
( g , kg , t )
E min
( g , kg , t )
n LC v min
( g , kg , t )
DR ( g , kg , t )

4.9.9  Definition of load cell families / construction

(This table is to be completed by the manufacturer for each load cell family within the model range)

Type / Model designationSpecificationOIML R 60-1Remark
Application of load3.2.1(e.g. tension / compression)
Load cell construction3.3(e.g. bending beam)
Material or combination of materials3.4.2
Shape3.4.2See R 60-2, 6.2.1
Design of measuring technique3.3.1(e.g. strain gauge bonded to metal)
Sealing of strain gauges3.4.2
Mounting methodAnnex E
Load transmissionAnnex ESee R 60-3, 4.9.1
Output rating3.4.2
Supply voltage3.4.2
Input impedance3.4.2
Cable connection3.4.2
Cable lengtha3.4.2

a  mandatory for strain gauge load cells with 4-wire connection

Further remarks concerning the definition of load cell families / construction (see table above)

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

4.9.10  Load cell dimensions within the load cell family

Pictures / Drawings of the load cell dimensions of the load cell family

4.9.11  Recommended load transmissions of the manufacturer

Pictures / Drawings of the recommended load transmissions

4.9.12  Results of previous tests that were taken into account

Model designationSerial numberJustification / RemarkTest Report No. (if available)

4.10  Information concerning the test equipment used for the tests

(including details of simulations and the way uncertainties are taken into account, including the level of “risk.” For instance, 95 % or k = 2 )

The following tables have to be completed for each individual piece of test equipment used for the tests.

General information:

For each of the following pieces of test equipment, indicate for which of the following test procedures the test equipment is used:

R 60 referenceTest procedure
R 60-2, 2.10.1Measurement error, repeatability error and temperature effect on minimum dead load output
R 60-2, 2.10.2Determination of creep error
R 60-2, 2.10.3Minimum dead load output return (DR)
R 60-2, 2.10.4Barometric pressure effects (atmospheric pressure)
R 60-2, 2.10.5Humidity effects for load cells marked with CH or no marked
R 60-2, 2.10.6Humidity effects for load cells marked SH
R 60-2, 2.10.7Additional tests for digital load cells

Example  A piece of test equipment is used for determination of the measurement error (R 60-2, 2.10.1), the creep error (R 60-2, 2.10.2), the minimum dead load output return (R 60-2, 2.10.3) and humidity effect marked with SH (R 60-2, 2.10.6):

R 60-2 reference2.10.12.10.22.10.32.10.42.10.52.10.52.10.62.10.7
Used for

4.10.1  Force generating system (if a force generating system or force generating machine is used)

DescriptionRemark
Designation
Type
Manufacturer
Identification number
Load range
Load steps
Unit
Preload
Rel. uncertainty ( k = 2 )
Last calibration
Certificate No. / report No.
Recalibration interval

The force generating system is used for the following test procedures:

R 60-2 reference2.10.12.10.22.10.32.10.42.10.52.10.52.10.62.10.7
Used for

Remarks / picture of the force generating system:

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

4.10.2  Weights

(if the load cell is tested manually with weights)

Number / identificationWeight
( g , kg , t )
Classa / rel. uncertainty ( k = 2 )Last calibrationRecalibration intervalCertificate No. / report No.

a  according to OIML R 111

The weights are used for the following test procedures:

R 60-2 reference2.10.12.10.22.10.32.10.42.10.52.10.52.10.62.10.7
Used for

Remarks / picture of the weights:

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

4.10.3  Temperature chamber (without humidity control)

DescriptionRemark
Designation
Type
Manufacturer
Identification number
Height × width × length dimension
Temperature range
Temperature stability
Humidity range
Humidity stability
Rel. uncertainty ( k = 2 )
Last calibration
Certificate No. / report No.
Recalibration interval

The climate chamber is used for the following test procedures:

R 60-2 reference2.10.12.10.22.10.32.10.42.10.52.10.52.10.62.10.7
Used for

Remarks / picture of the temperature chamber:

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

4.10.4  Climate chamber (with temperature and humidity control)

DescriptionRemark
Designation
Type
Manufacturer
Identification number
height × width × length dimension
Temperature range
Temperature stability
Humidity range
Humidity stability
Rel. uncertainty ( k = 2 )
Last calibration
Certificate No. / report No.
Recalibration interval

The climate chamber is used for the following test procedures:

R 60-2 reference2.10.12.10.22.10.32.10.42.10.52.10.52.10.62.10.7
Used for

Remarks / picture of the climate chamber:

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

4.10.5  Indicator / Indicating instrument

(for testing analogue load cells)

DescriptionRemark
Designation
Type
Manufacturer
Identification / Serial number
Measurement range
Rel. uncertainty ( k = 2 )
Last calibration
Certificate No. / report No.
Recalibration interval

Settings of the indicator / indicating instrument used for the tests

DescriptionRemark
Measurement range
Supply voltage (AC/DC)
Filter settings
Cable connections

The indicator / indicating instrument is used for the following test procedures:

R 60-2 reference2.10.12.10.22.10.32.10.42.10.52.10.52.10.62.10.7
Used for

Remarks / picture of the indicator / indicating instrument:

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

4.10.6  Terminal / Digital data processing device

(for testing digital load cells)

DescriptionRemark
Designation
Type
Manufacturer
Identification / Serial number
Measurement range
Last calibration
Certificate No. / report No.
Recalibration interval

Settings of the indicator / indicating instrument used for the tests

DescriptionRemark
Measurement range
Supply voltage (AC/DC)
Filter settings
Cable connections

The terminal / digital data processing device is used for the following test procedures:

R 60-2 reference2.10.12.10.22.10.32.10.42.10.52.10.52.10.62.10.7
Used for

Remarks / picture of the terminal / digital data processing device:

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

4.10.7  Barometric pressure meter

DescriptionRemark
Type
Manufacturer
Identification / Serial number
Measurement range
Rel. uncertainty ( k = 2 )
Last calibration
Certificate No. / report No.
Recalibration interval

The barometric pressure meter is used for the following test procedures:

R 60-2 reference2.10.12.10.22.10.32.10.42.10.52.10.52.10.62.10.7
Used for

4.10.8  Thermometer

DescriptionRemark
Type
Manufacturer
Identification / Serial number
Measurement range
Rel. uncertainty ( k = 2 )
Last calibration
Certificate No. / report No.
Recalibration interval

The thermometer is used for the following test procedures:

R 60-2 reference2.10.12.10.22.10.32.10.42.10.52.10.52.10.62.10.7
Used for

4.10.9  Moisture analyser

DescriptionRemark
Type
Manufacturer
Identification / Serial number
Measurement range
Rel. uncertainty ( k = 2 )
Last calibration
Certificate No. / report No.
Recalibration interval

The moisture analyser is used for the following test procedures

R 60-2 reference2.10.12.10.22.10.32.10.42.10.52.10.52.10.62.10.7
Used for

4.10.10  Additional test equipment

(e.g. burst generator for testing of digital load cells)

DescriptionRemark
Test equipment
Type
Manufacturer
Identification / Serial number
Measurement range
Rel. uncertainty ( k = 2 )
Last calibration
Certificate No. / report No.
Recalibration interval

The equipment is used for the following test procedures:

R 60-2 reference2.10.12.10.22.10.32.10.42.10.52.10.52.10.62.10.7
Used for

4.10.11  Remarks (settings, pictures, further information)

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

5  Examination

(To be completed by the Evaluating Authority)

5.1  Marking requirements (R 60-1, 6.2)

5.1.1  Mandatory markings on the load cell (R 60-1, 6.2.1)

R 60-1 referenceInformationFulfils requirements
YesNo
6.2.1Name or trademark of manufacturer
6.2.1Manufacturer’s own designation or load cell model
6.2.1Serial number
6.2.1Maximum capacity, E max a
6.2.1Year of production
6.2.1Type approval mark (if applicable)

a  In units of ( g , kg , t )

5.1.2  Mandatory markings on the load cell or an accompanying document

(R 60-1, 6.2.2)

R 60-1 referenceMandatory informationOn load cellIn documentFulfils requirements
YesNo
6.2.4.1Accuracy classes and their symbols
6.2.4.5Maximum number of load cell verification intervals, n LC
6.2.4.2Loading designation (if necessary)
6.2.4.3Working temperature designation
6.2.4.4Humidity symbol “NH”
6.2.4.4Humidity symbol “SH”
6.2.2Minimum dead load, E min
6.2.2Safe load limit, E lim
5.1.3, 6.2.2Minimum load cell verification interval ( v min )
6.2.2Other pertinent conditions
3.7.2, 5.3.2Apportionment factor, p LC (if not equal to 0.7)
5.1.6Standard classification
5.1.7Multiple classifications

5.1.3  Non-mandatory, additional information (R 60-1, 6.2.3)

R 60-1 referenceMandatory informationOn load cellIn documentFulfils requirements
YesNo
5.6.3.1Humidity symbol “CH”
3.5.15Relative v min , Y
3.5.14Relative DR , Z

5.2  Suitability for testing (R 60-2, 2.3, 2.4)

Date:

Observer:

Serial number:

Fulfils requirements

Yes

No

Remarks

Passed ☐ Yes ☐ No

5.3  Software (if present) (R 60-1, 6.1)

Date:

Observer:

Serial number:

Version of software:

Identification code:

Yes

No

Software protected by sealing

Automatic change of identification code

Fixed version number

Remarks: ‌

Passed ☐ Yes ☐ No

5.4  Documentation for type approval (R 60-2, 2.5)

YesNoRemarks

a) Description of the general principle of measurement (R 60-2, 2.5, a)

b) List and characteristics of essential components + details

c) Mechanical drawings (R 60-2, 2.5, b)

d) Electric/electronic diagrams (R 60-2, 2.5, c)

e) Installation requirements (R 60-2, 2.5, d)

f) Sealing plan

g) Panel layout

h) General information of the software (R 60-2, 2.5, g)

For details, see R 60-1, 6.1

  1. Operating instructions (R 60-2, 2.5, e)

j) Information supporting the manufacturer’s assumption of compliance (R 60-2, 2.5, f)

Other relevant information pertaining to identification of the instrument, diagrams, results of previous tests, etc.: (attach photograph(s) and/or outline-drawing(s) here if available):

Remarks:

Passed

☐ Yes ☐ No

6  Performance tests

6.1  Results of the performance tests

Clause R 60-1/2Performance testsTemperature in °C report page No.Maximum error in vPassedFailedRemark
R 60-1, 5.3 / R 60-2, 2.10.1Load cell errors ( E L ) (see R 60-3, 2.1.2)
R 60-1, 5.4 / R 60-2, 2.10.1Repeatability errors ( E R ) (see R 60-3, 2.1.3)
R 60-1, 5.5.1 / R 60-2, 2.10.2Creep ( C C ( t ) ) (see R 60-3, 2.1.5)
R 60-1, 5.5.1 / R 60-2, 2.10.2Creep ( C C ( 30 20 ) ) (see R 60-3, 2.1.5.2)
R 60-1, 5.5.2 / R 60-2, 2.10.3Minimum dead load output return ( C DR ) (see R 60-3, 2.1.5.4)(See Note 1)
DR =
(See Note 1)
DR =
(See Note 1)
DR =
(See Note 1)
DR =
R 60-1, 5.6.3.1 / R 60-2, 2.10.5Humidity effects ( C H min )
(CH or no mark)
(see R 60-3, 2.1.7.1)
R 60-1, 5.6.3.1 / R 60-2, 2.10.5Humidity effects ( C H max )
(CH or no mark)
(see R 60-3, 2.1.7)
R 60-1, 5.6.3.2 / R 60-2, 2.10.6Humidity effects (SH)
(see R 60-3, 2.1.8)
R 60-1, 5.6.1.3 / R 60-2, 2.10.1Temperature effects on
minimum dead load output
( C M )
(see R 60-3, 2.1.4)
(See Note 2)
R 60-1, 5.6.2 / R 60-2, 2.10.4Barometric pressure
effects ( C P ( v min ) )
(see R 60-3, 2.1.6)
(See Note 2)

Note 1  DR is the minimum dead load output return in units of ( g , kg , t ) and determined according to R 60-3, 2.1.5.8

Note 2  Maximum error in unit v min

Remarks:

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

6.1.1  Results of the Performance tests for digital load cells

Clause R 60-1/2Performance testsTemperature in °C report page No.Maximum error in vPassedFailedRemark
R 60-1, 5.7.2.1 / R 60-2, 2.10.7.3Warm-up time
(see R 60-3, 2.2.1)
R 60-1, 5.7.2 / R 60-2, 2.10.7.4Power voltage variations
(see R 60-3, 2.2.2)
R 60-1, 5.7.2.5 / R 60-2, 2.10.7.5Short time power reductions
(see R 60-3, 2.2.3)
R 60-1, 5.7.2.5 / R 60-2, 2.10.7.6Bursts (electrical fast transients)
(see R 60-3, 2.2.4)
R 60-1, 5.7.2.5 / R 60-2, 2.10.7.7Surge (see R 60-3, 2.2.5)
R 60-1, 5.7.2.5 / R 60-2, 2.10.7.8Electrostatic discharge
(see R 60-3, 2.2.6)
R 60-1, 5.7.2.5 / R 60-2, 2.10.7.9Electromagnetic susceptibility
(see R 60-3, 2.2.7)
R 60-1, 5.7.2.5 / R 60-2, 2.10.7.10Immunity to conducted electromagnetic fields
(see R 60-3, 2.2.8)
R 60-1, 5.7.2.6 / R 60-2, 2.10.7.11Span stability
(see R 60-3, 2.2.9)

Remarks:

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

_________________________________________________________________________________

6.2  Initial tests and general notes concerning performance tests

(To be completed or under the responsibility of the Evaluating Authority)

6.2.1  Units

Unit (e.g. counts, digits, g, kg, t) in which the measurement result is displayed.

R 60-2 referenceTest procedureUnit
2.10.1Measurement error, repeatability error and temperature effect on minimum dead load output
2.10.2Determination of creep error
2.10.3Minimum dead load output return (DR)
2.10.4Barometric pressure effects (Atmospheric pressure)
2.10.5Humidity effects for load cells marked with CH or no marked
2.10.6Humidity effects for load cells marked SH
2.10.7Additional tests for analogue-active cells

6.2.2  Measurement range (R 60-1, 5.2, 5.5.2)

Fulfils requirements

Test procedure
(R 60-2 reference)

D max

D min

Conversion factor, f [ indication / v ]
(see R 60-3, 2.1.2.4)

Yes

No

2.10.1

2.10.2

2.10.3

2.10.4

2.10.5

2.10.6

2.10.7

Passed

☐Yes

☐No

6.2.3  Conditions

(see R 60-2, 2.8.1)

(To ensure that these requirements are met, the calculations should be carried out using lower n values than the n LC specified. The calculations made do not include the application of 2.8.1).

Check that

v min D max D min n

It should be sufficient to carry out the calculations with n = n LC , n max 500 and n = n LC 1 000 if applicable.

Test procedure (R 60-2 reference) D min D max n LC Is the requirement v min D max D min n fulfilled with
n LC n LC 500 n LC 1 000
YesNoYesNoYesNo
2.10.1
2.10.2
2.10.3
2.10.4
2.10.5
2.10.6
2.10.7

Passed ☐ Yes ☐ No

6.2.4  Input impedance

Measure the input impedance and compare the result with the input impedance in OIML R 60-3, 4.5.

6.3  Load test data (Load cell error E L ) 3 runs

Ref.: R 60-2, 2.10.1.1 to 2.10.1.11. Complete one sheet for each test temperature, one for each humidity (SH) test in 2.10.6, and when applicable, one for each electronics power voltage in 2.10.7.4.

Application no.:

Load cell model:

Serial no.:

E max :

n LC :

v min :

p LC : DR:

Force-generating system:

Indicating instrument:

Evaluator:

At startAt end
Date:
Temperature:

°C

Relative humidity:

%

Barometric pressure:

kPa

Indicator temperature:

°C

Electronics power voltage (when applicable):

Table 1 — (3 runs)
Test load
(units)
Run no. 1Run no. 2Run no. 3Average indication
(counts)
Repeatability error
(counts)
Indication
(counts)
Time
(hh mm ss)
Indication
(counts)
Time
(hh mm ss)
Indication
(counts)
Time
(hh mm ss)
0
0
0
0
0*

Note 

  1. * = Average initial minimum test load indication.

  2. Absolute (not relative) time shall be recorded.


6.4  Load test data (Load cell error E L ) 5 runs

R 60-2, 2.10.1.1 to 2.10.1.11. Complete one sheet for each test temperature, one for each humidity (SH) test in 2.10.6, and when applicable, one for each electronics power voltage in 2.10.7.4.

Application no.:

Load cell model:

Serial no.:

E max :

n LC :

v min :

p LC : DR:

Force-generating system:

Indicating instrument:

Evaluator:

At startAt end
Date:
Temperature:

°C

Relative humidity:

%

Barometric pressure:

kPa

Indicator temperature:

°C

Electronics power voltage (when applicable):

Table 2 — (5 runs)
Test load (units)Run no. 1Run no. 2Run no. 3Run no. 4Run no. 5Average indication (counts)Repeatability error (counts)
Indication
(counts)
Time
(hh mm ss)
Indication
(counts)
Time
(hh mm ss)
Indication
(counts)
Time
(hh mm ss)
Indication
(counts)
Time
(hh mm ss)
Indication
(counts)
Time
(hh mm ss)
0
0
0
0
0*

Note 

  1. * = Average initial minimum test load indication.

  2. Absolute (not relative) time shall be recorded.


6.5  Load cell errors ( E L ) calculation

R 60-1, 5.3.1
R 60-2, 2.10.1.12 to 2.10.1.14
R 60-3, 2.1.2.2

Application no.:

Load cell model:

Serial no.:

E max :

n LC :

v min :

p LC : DR:

Force-generating system:

Indicating instrument:

Evaluator:

At startAt end
Date:
Temperature:

°C

Relative humidity:

%

Barometric pressure:

kPa

Indicator temperature:

°C

Conversion factor, f :

75 % test load ( g , kg , t ):

Reference indication at 75 % test load:

Table 3
Test load (units)Reference indication (counts). . . . °C ( 20 °C ). . . . °C ( T 1 °C ). . . . °C ( T 2 °C ). . . . °C ( 20 °C )MPE
( v )
Indications (counts)Error ( E L ) ( v )Indication (counts)Error ( E L ) ( v )Indication (counts)Error ( E L ) ( v )Indication (counts)Error ( E L ) ( v )
000000

Minimum test load, D min : PASS: FAIL:

Note 

  1. Load/reference indications: if a 75 % load point was not obtained, a straight line interpolation between the adjacent higher and lower load point indications is used (see R 60-1, 5.3.1 and calculation procedures in R 60-3, 2.1.2.2).

  2. Error, E L : the difference between the test indication and the reference indication divided by the conversion factor, f .

  3. Test load values are values above minimum test load, D min .

6.6  Repeatability errors ( E R ) calculation

R 60-1, 5.4
R 60-2, 2.10.1.15
R 60-3, 2.1.3

Application no.:

Load cell model:

Serial no.:

E max :

n LC :

v min :

p LC : DR:

Force-generating system:

Indicating instrument:

Evaluator:

Conversion factor, f :

Table 4
Test load (unit)…​. °C ( 20 °C )…​. °C ( T 1 °C )…​. °C ( T 2 °C )…​. °C ( 20 °C )MPE ( v )

Repeatability error (counts)

Repeatability error v

Repeatability error (counts)

Repeatability error v

Repeatability error (counts)

Repeatability error v

Repeatability error (counts)

Repeatability error v

PASS: FAIL:

Note  Error, E R : the maximum difference between the three test indications divided by the conversion factor, f (classes C and D) or the maximum difference between the five test indications divided by the conversion factor, f (classes A and B).

6.7  Temperature effects on minimum dead load output (MDLO)

R 60-1, 5.5.2
R 60-2, 2.10.1.16
R 60-3, 2.1.4

Application no.:

Load cell model:

Serial no.:

E max :

n LC :

v min :

p LC : DR:

Force-generating system:

Indicating instrument:

Evaluator:

Conversion factor, f :

Table 5
Temprature
( °C )
Indication
(     )
Change
( v )
Change
( v min / . . . °C )
mpc
( v min / . . . °C )
p LC
p LC
p LC

PASS: ☐ FAIL: ☐

Note 

  1. MDLO: minimum dead load output.

  2. Indication: the average initial minimum test load indication obtained from Table 1.

  3. The maximum permissible change (mpc) allowed is: ( v min / 5 °C ) for classes B, C, and D; ( v min / 2 °C ) for class A.

  4. Change, C M ( v ): the difference between the observed indications, and the indications at the prior temperature, divided by the conversion factor, f .

6.8  Creep ( C C ) and DR ( C DR )

R 60-1, 5.5.1, 5.5.2
R 60-2, 2.10.2, 2.10.3. Complete one sheet for each test temperature.

Application no.:

Load cell model:

Serial no.:

E max :

n LC :

v min :

p LC : DR:

Force-generating system:

Indicating instrument:

Evaluator:

At startAt end
Date:
Temperature:

°C

Relative humidity:

%

Barometric pressure:

kPa

Indicator temperature:

°C

Conversion factor, f :

Table 6
CreepDR

Test load

Original

Barom. Press.

Change of

mpc

Test load

Original

Barom. Press.

Change of

mpc

Indication

Time

Indication

Time

Indication

Time

Indication

Time

counts

hh:mm:ss

kPa

v

mm:ss

v

counts

hh:mm:ss

kPa

v

mm:ss

v

D min

D max

(*)

D max

D min

(***)

D max

D min

(**)

DR ( v ): 30 minute creep: PASS: FAIL:
actual time (s): 20 — 30 minute creep difference ( < 0,15 MPE ): PASS: FAIL:
specified time (s): DR < 0.5 v : PASS: FAIL:
MPE for DR ( v ): DR within manuf. Specified DR requirements: PASS: FAIL:

Note 

  1. Change ( v ) for creep: the observed indication minus the initial “load” indication (**) divided by the conversion factor, f .

  2. Determine the difference between the reading obtained at 20 minutes and the reading obtained at 30 minutes (see 5.5.1).

  3. Change ( v ) for DR: the initial indication (***) minus the initial “no load” indication (*) divided by the conversion factor, f .

  4. Absolute (not relative) time shall be recorded.

6.9  Barometric pressure effects ( C P )

R 60-1, 5.6.2
R 60-2, 2.7.3.8, 2.10.4
R 60-3, 2.1.6

Complete one sheet for each test temperature.

Application no.:

Load cell model:

Serial no.:

E max : E min :

n LC : p LC :

Y : Z :

v min : DR:

Force-generating system:

Test load:, D max : D min :

Indicating instrument:

Evaluator:

At startAt end
Date:
Temperature:

°C

Relative humidity:

%

Barometric pressure:

kPa

Indicator temperature:

°C

Conversion factor, f :

Table 7
Pressure
( kPa )
Indication
(counts)
Time
hh:mm
Change
( v )
Change
( v min / kPa )
mpc
( v min / kPa )
000
1

PASS: ☐ FAIL: ☐

6.10  Humidity effects

6.10.1  Humidity effects (CH or no mark)

R 60-1, 5.6.3
60-2, 2.7.3.9, 2.10.5
R 60-3, 2.1.7

6.10.1.1  Form 6.10.1.(a): Humidity effects summary (CH or no mark)

Application no.:

Load cell model:

Serial no.:

E max : E min :

n LC : p LC :

Y : Z :

v min : DR:

Force-generating system:

Test load:, D max : D min :

Indicating instrument:

Evaluator:

At startAt end
Date:
Temperature:

°C

Relative humidity:

%

Barometric pressure:

kPa

Indicator temperature:

°C

Conversion factor, f :

Table 8
Test load
(g, kg, or t)
Before humidity testAfter humidity testChange ( v )mpc ( v )
Indication
(counts)
Time
(hh mm ss)
Indication
(counts)
Time
(hh mm ss)
Average indication at D min (¤)           C Hmin = < 4 % n LC
Average indication at D max (‡)
Average difference (*)           C Hmax = 1

(¤) Indications at minimum test load

Change (¤), C Hmin : PASS: ☐ FAIL: ☐

(‡) Indications at maximum test load c)

(*) Average, see R 60-1, 5.6.3 and R 60-3, 2.1.7

Change (*), C Hmax : PASS: ☐ FAIL: ☐

Note 

  1. This test is not necessary if the load cell is marked NH or SH.

  2. Change ( v ): the difference between the indication after and before humidity exposure divided by the conversion factor, f .

  3. Use five test runs for Class A and B; use 3 test runs for Class C and D.

  4. Absolute (not relative) time shall be recorded.

  5. For family certification this test is not necessary, if a pattern with a smaller capacity and the same or better metrological characteristics has passed this test.

6.10.2  Form 6.10.1.(b): Load test data ( E L ) — 3 runs

R 60-2, 2.10.1.1-2.10.1.11. Complete this form if the measurement error is determined before the humidity test (CH) is carried out (not mandatory)

Application no.:

Load cell model:

Serial no.:

E max : E min :

n LC : p LC :

Y : Z :

v min : DR:

Force-generating system:

Test load:, D max : D min :

Indicating instrument:

Evaluator:

At startAt end
Date:
Temperature:

°C

Relative humidity:

%

Barometric pressure:

kPa

Indicator temperature:

°C

Electronics power voltage (when applicable):

Table 9 — 3 runs
Test load
(unit)
Run no. 1Run no. 2Run no. 3Average indication countsRepeatability error counts
Indication
counts
Time
hh:mm:ss
Indication
counts
Time
hh:mm:ss
Indication
counts
Time hh:mm:ss
*

Note  * Average initial minimum test load indication

Absolute (not relative) time shall be recorded

6.10.2.1  Form 6.10.1.(c): Load test data ( E L ) — 3 runs

R 60-2, 2.10.1.1-2.10.1.11. Complete this form if the measurement error is determined after the humidity test (CH) is carried out (not mandatory)

Application no.:

Load cell model:

Serial no.:

E max : E min :

n LC : p LC :

Y : Z :

v min : DR:

Force-generating system:

Test load:, D max : D min :

Indicating instrument:

Evaluator:

At startAt end
Date:
Temperature:

°C

Relative humidity:

%

Barometric pressure:

kPa

Indicator temperature:

°C

Electronics power voltage (when applicable):

Table 10 — (3 runs)
Test load (unit)Run no. 1Run no. 2Run no. 3Average indication countsRepeatability error counts
Indication
counts
Time
hh:mm:ss
Indication
counts
Time
hh:mm:ss
Indication
counts
Time
hh:mm:ss
*

Note  * Average initial minimum test load indication

Absolute (not relative) time shall be recorded


6.10.3  Form 6.10.1.(d): Load test data ( E L ) — 5 runs

R 60-2, 2.10.1.1-2.10.1.11. Complete this form if the measurement error is determined before the humidity test (CH) is carried out (not mandatory)

Application no.:

Load cell model:

Serial no.:

E max :

n LC :

v min :

p LC : DR:

Force-generating system:

Indicating instrument:

Evaluator:

At startAt end
Date:
Temperature:

°C

Relative humidity:

%

Barometric pressure:

kPa

Indicator temperature:

°C

Electronics power voltage (when applicable):

Table 11 — Table 6.10.1.(d) 5 runs
Test load (unit)Run no. 1Run no. 2Run no. 3Run no. 4Run no. 5Average indication countsRepeatability error counts
Indication countsTime hh:mm:ssIndication countsTime hh:mm:ssIndication countsTime hh:mm:ssIndication countsTime hh:mm:ssIndication countsTime hh:mm:ss
*

Note  * Average initial minimum test load indication

6.10.4  Form 6.10.1.(e):Load test data ( E L ) — 5 runs

R 60-2, 2.10.1.1-2.10.1.11. Complete this form if the measurement error is determined after the humidity test (CH) is carried out (not mandatory)

Application no.:

Load cell model:

Serial no.:

E max : E min :

n LC : p LC :

Y : Z :

v min : DR:

Force-generating system:

Test load:, D max : D min :

Indicating instrument:

Evaluator:

At startAt end
Date:
Temperature:

°C

Relative humidity:

%

Barometric pressure:

kPa

Indicator temperature:

°C

Electronics power voltage (when applicable):

Table 12 — 5 runs
Test load (unit)Run no. 1Run no. 2Run no. 3Run no. 4Run no. 5Average indication countsRepeatability error counts
Indication countsTime hh:mm:ssIndication countsTime hh:mm:ssIndication countsTime hh:mm:ssIndication countsTime hh:mm:ssIndication countsTime hh:mm:ss
*

Note  * Average initial minimum test load indication


6.10.5  Humidity effects (SH)

6.10.5.1  Form 6.10.2 Humidity effects (SH) summary

R 60-1, 5.6.3.2
R 60-2, 2.7.3.9, 2.10.6
R 60-3, 2.1.8

Application no.:

Load cell model:

Serial no.:

E max :

n LC :

v min :

p LC : DR:

Force-generating system:

Indicating instrument:

Evaluator:

At startAt end
Date:
Conditioning period:
Reference temperature:

°C

High temperature:

°C

Reference relative humidity:

%

High relative humidity:

%

Conversion factor, f :

Page of load test before humidity test:

Page of load test during humidity test:

Page of load test after humidity test:

For summary of SH-humidity load test errors: use form 6.3 (3 runs) or 6.4 (5 runs) as appropriate to record individual teat results.

Table 13
Test load kg Reference indication (counts)…. °C ( 20 °C ) …. % (50 %) RH…. °C (High) …. % (85 %) RH…. °C ( 20 °C ) …. % (50 %) RHMPE v
Indication (counts)Error ( E L ) v Indication (counts)Error ( E L ) v Indication (counts)Error ( E L ) v

PASS: FAIL:

Note 

  1. Load/reference indications: if at 75 % load point was not obtained, a straight line interpolation between the adjacent higher and lower load point indication is used.

  2. Error, E L : the difference between the test reference and the reference indication divided by the conversion factor, f .

  3. Test load values are values above minimum test load, D min .

  4. Conditioning period: the time period for exercising the load cell.

  5. For family certification this test is not necessary, if a pattern with a smaller capacity and the same or better metrological characteristics has passed this test.

6.11  Warm-up time

6.11.1  Form 6.11 Warm-up time

R 60-1, 3.5.17
R 60-2, 2.10.7.3
R 60-3, 2.2.1

Application no.:

Load cell model:

Serial no.:

E max : E min :

n LC : p LC :

Y : Z :

v min : DR:

Force-generating system:

Test load:, D max : D min :

Indicating instrument:

Evaluator:

At startAt end
Date:
Temperature:

°C

Relative humidity:

%

Barometric pressure:

kPa

Indicator temperature:

°C

Conversion factor, f : counts/ v

Duration of disconnection before test:

Table 14
Initial runAfter 5 min.After 15 min.After 30 min.mpc v min
Indication (counts)Time hh:mm:ssIndication (counts)Time hh:mm:ssIndication (counts)Time hh:mm:ssIndication (counts)Time hh:mm:ss
D min
D max
SpanCounts
Span v min
Change v min

PASS: ☐ FAIL: ☐

Note 

  1. Absolute (not relative) time shall be recorded.

  2. Span: the result of subtraction of the indication at minimum test load from the indication at maximum test load. All span errors (error at maximum test load minus the error at minimum test load) shall be within the maximum permissible error during the 30 minute test.

  3. The change of span must not exceed v min .

  4. Change: the difference between the span and the initial run span.

  5. Maximum permissible change, mpc: the absolute value of the maximum permissible error for the maximum test load applied.

  6. Exercises have to be run before disconnection.

6.12  Power voltage variation

6.12.1  Form 6.12 Power voltage variation

R 60-1, 5.7.2.2, 5.7.2.3, 5.7.2.4
R 60-2, 2.10.7.4
R 60-3, 2.2.2

Application no.:

Load cell model:

Serial no.:

E max : E min :

n LC : p LC :

Y : Z :

v min : DR:

Force-generating system:

Test load:, D max : D min :

Indicating instrument:

Evaluator:

At startAt end
Date:
Temperature:

°C

Relative humidity:

%

Barometric pressure:

kPa

Indicator temperature:

°C

Conversion factor, f : counts/ v

Main voltage:

AC: DC:

Table 15
Test load (units)Preloads
Indication (counts)Time hh:mm:ss
D min
D max
D min
D max
D min
D max

Note 

  1. Reference indications: if at 75 % load point was not obtained, a straight line interpolation between the adjacent higher and lower indication is used (see 2.8.2 in R 60-2 and calculation procedures in R 60-3, 2.1.2)

  2. Error: the difference between the test indication and the reference indication divided by the conversion factor, f .

  3. The change of span must not exceed v min .

  4. When a voltage range is marked, use the average value as the reference value and determine upper and lower values of applied voltage according to R 60-2, 2.10.7.4.

  5. Upper limit not applicable to battery powered load cells

  6. At lower limit, battery powered load cells shall function and be within MPE, or cease to function

Initial run with main voltagelower limit main voltage — 15 %upper limit +10 %mpc
v min
Indication (counts)Time hh:mm:ssIndication (counts)Time hh:mm:ssIndication (counts)Time hh:mm:ss
D min
D max
SpanCounts
Span v min
Change v min

PASS: ☐ FAIL:☐

6.13  Short time power reductions

6.13.1  Form 6.13 Short time power reductions

R 60-1, 5.7.2.5
R 60-2, 2.10.7.5
R 60-3, 2.2.3

Application no.:

Load cell model:

Serial no.:

E max :

n LC :

v min :

p LC : DR:

Force-generating system:

Indicating instrument:

Evaluator:

Date:

Time:

Temperature: °C

Relative humidity: %

Barometric pressure: kPa

Conversion factor, f :

Minimum test load, D min :

Reference voltage range: V

Table 16
Test load ( g , kg , t )DisturbanceResult
Amplitude (%)Duration (cycles)Number of disturbancesRepetition interval ( v )Indication
(   )
Difference
(   )
Significant fault > v min
NoYes (remarks)
Without disturbance
00.510
50110

PASS: ☐ FAIL: ☐

Equipment used (supply sketch if necessary):

Remarks:

Note  In the case of a voltage range, use the average value as the reference value.

6.14  Bursts (electrical fast transients)

6.14.1  Form 6.14.1 Bursts (electrical fast transients) — power supply lines

R 60-1, 5.7.2.5
R 60-2, 2.10.7.6
R 60-3, 2.2.4

Application no.:

Load cell model:

Serial no.:

E max :

n LC :

v min :

p LC : DR:

Force-generating system:

Indicating instrument:

Evaluator:

Date:

Time:

Temperature: °C

Relative humidity: %

Barometric pressure: kPa

Conversion factor, f :

Minimum test load, D min :

Table 17 — Power supply lines: test voltage = 2 kV; duration of test = 1 minute at each polarity
Test load (g, kg, or t)ConnectionPolarityResult
L to groundN to groundPE to groundIndication
(     )
Difference
( v )
Significant fault > v min
NoYes (remarks)
without disturbance
xpos
neg
without disturbance
xpos
neg
without disturbance
xpos
neg

PASS: ☐ FAIL: ☐

L = phase, N = neutral, PE = protective earth

Equipment used (supply sketch if necessary)

6.14.2  Form 6.14.2 Bursts (electrical fast transients) — I/O circuits and communications lines

R 60-1, 5.7.2.5
R 60-2, 2.10.7.6
R 60-3, 2.2.4

Application no.:

Load cell model:

Serial no.:

E max :

n LC :

v min :

p LC : DR:

Force-generating system:

Indicating instrument:

Evaluator:

Date:

Time:

Temperature: °C

Relative humidity: %

Barometric pressure: kPa

Conversion factor, f :

Minimum test load, D min :

Table 18
Test load (g, kg, or t)Cable interfacePolarityResult
Indication
(   )
Difference
( v )
Significant fault > v min
NoYes (remarks)
without disturbance
pos
neg
without disturbance
pos
neg
without disturbance
pos
neg
without disturbance
pos
neg
without disturbance
pos
neg
without disturbance
pos
neg

PASS: ☐ FAIL: ☐

Equipment used (supply sketch if necessary)

Remarks:

Note  Explain or make a sketch indicating where the clamp is located on the cable: if necessary use additional page(s).

6.15  Surges

6.15.1  Form 6.15 Surges

R 60-1, 5.7.2.5
R 60-2, 2.10.7.7
R 60-3, 2.2.5

Application no.:

Load cell model:

Serial no.:

E max :

n LC :

v min :

p LC : DR:

Force-generating system:

Indicating instrument:

Evaluator:

Conversion factor, f :

Minimum test load, D min :

Table 19

OIML
R 60-2,
…​…​…​

[unit]
☐[g];
☐[kg];
☐[t];

Test conditions surges on signal, data and control lines

Observer’s name:

Output gained

using actual loads

Test load:

Line to line 1 kV

simulating loading

Line to earth 2 kV

using:

Cable:

☐ Symmetrical line

Date:

Start

Stop

☐ Unsymmetrical line

Time:

Specimen:

Ambient temperature

°C

°C

f

Relative humidity

%

%

D min [unit]

Barometric Pressure

kPa

kPa

D max [unit]

Cycle phase

Initial

During exposure

After

Load

Time

Start

Stop

Quantity
[unit]

reference

indicated

Error [ v min ]

relative error [%] E ii

MPE [%]

Pass

Fail

Observed faults after exposure

Fault limit [%]

………….

Line to line (N/A for balanced)

Fault/Deviation

Significant

Acts on fault

↑ ▇

↓ ▇

Yes

No

Yes

No

3x

3x

Line to earth

3x

3x

Observations

Result

Pass

Fail

6.16  Electrostatic discharge

6.16.1  Form 6.16.1 Electrostatic discharge — direct application

R 60-1, 5.7.2.5
R 60-2, 2.10.7.8
R 60-3, 2.2.6

Application no.:

Load cell model:

Serial no.:

E max :

n LC :

v min :

p LC : DR:

Force-generating system:

Indicating instrument:

Evaluator:

Date:

Time:

Temperature: °C

Relative humidity: %

Barometric pressure: kPa

Conversion factor, f :

Minimum test load, D min :

Contact discharges

Paint penetration

Air discharges

Polarity (see Note 2):

Positive

Negative

Table 20
Test load ( g , kg , t )Result
Test voltage
( kV )
No. of discharges
10
Repetition interval
( s )
Indication
(   )
Difference
( v )
Significant fault > v min
NoYes (remarks)
without disturbance
2
4
6
8 (air discharges)

PASS: ☐ FAIL: ☐

Remarks:

Note 1  If the load cell fails, the test point at which this occurs shall be recorded.

Note 2  IEC Publication 61000-4-2 Ed 2.0 (2008-12) Consolidated edition specifies that the test be conducted with the most sensitive polarity.

6.16.2  Form 6.16.2 Electrostatic discharge — indirect application

R 60-1, 5.7.2.5
R 60-2, 2.10.7.8
R 60-3, 2.2.6

Application no.:

Load cell model:

Serial no.:

E max :

n LC :

v min :

p LC : DR:

Force-generating system:

Indicating instrument:

Evaluator:

Date:

Time:

Temperature: °C

Relative humidity: %

Barometric pressure: kPa

Conversion factor, f :

Minimum test load, D min :

Polarity (see Note 2):

Positive

Negative

Table 21 — Horizontal coupling plane
Test load ( g , kg , t )Result
Test voltage
( kV )
No. of discharges
10
Repetition interval
( s )
Indication
(   )
Difference
( v )
Significant fault > v min
NoYes (remarks)
without disturbance
2
4
6
Table 22 — Vertical coupling plane
Test load ( g , kg , t )Result
Test voltage
( kV )
No. of discharges
10
Repetition interval
( s )
Indication
(   )
Difference
( v )
Significant fault > v min
NoYes (remarks)
without disturbance
2
4
6

PASS: ☐ FAIL: ☐

Remarks:

Note 1  If the load cell fails, the test point at which this occurs shall be recorded.

Note 2  IEC Publication 61000-4-2 Ed. 2.0 (2008-12). Consolidated edition specifies that the test be conducted with the most sensitive polarity.

6.16.3  Form 6.16.3 Electronic discharge (continued) — specification of test points

R 60-1, 5.7.2.5
R 60-2, 2.10.7.8
R 60-3, 2.2.6

Specify test points utilised on load cell and test equipment used, e.g. by photos or sketches.

  1. Direct application

    Contact discharges:

    Air discharges:

  2. Indirect application

6.17  Electromagnetic susceptibility

6.17.1  Form 6.17.1 Electromagnetic susceptibility

R 60-1, 5.7.2.5
R 60-2, 2.10.7.9
R 60-3, 2.2.7

Application no.:

Load cell model:

Serial no.:

E max :

n LC :

v min :

p LC : DR:

Force-generating system:

Indicating instrument:

Evaluator:

Date:

Time:

Temperature: °C

Relative humidity: %

Barometric pressure: kPa

Conversion factor, f :

Minimum test load, D min :

Rate of sweep:

Test load:

Test load material:

Table 23
DisturbanceResult
AntennaFrequency range (MHz)PolarisationFacing load cellIndication
(   )
Difference
( v )
Significant fault > v min
NoYes (remarks)
without disturbance
VerticalFront
Right
Left
Rear
HorizontalFront
Right
Left
Rear

PASS: ☐ FAIL: ☐

Frequency range: 80 MHz 4) to 3 000 MHz

Field strength: 10 V m 1

Modulation: 80 % AM, 1 kHz sine wave

Remarks:

Note  If the load cell fails, the test point at which this occurs shall be recorded.

6.17.2  Form 6.17.2 Electromagnetic susceptibility (continued) — description of the test setup

Describe the setup of the test and equipment, e.g. by photos or sketches:


6.18  Immunity to conducted electromagnetic fields

6.18.1  Form 6.18 Immunity to conducted electromagnetic fields

R 60-1, 5.7.2.5
R 60-2, 2.10.7.10
R 60-3, 2.2.8

Application no.:

Load cell model:

Serial no.:

E max :

n LC :

v min :

p LC : DR:

Force-generating system:

Indicating instrument:

Evaluator:

Date:

Time:

Temperature: °C

Relative humidity: %

Barometric pressure: kPa

Conversion factor, f :

Minimum test load, D min :

Rate of sweep:

Test load:

Test load material:

Table 24

OIML
R 60-2,
…​…​..

[unit]
☐ [g];
☐ [kg];
☐ [t]

test conditions rf current injection

Observer’s name:

output gained

using actual loads

Test load:

f l =          MHz

simulating loading

f h =          MHz

using:

RF voltage v emf

Cable exposed

Modulation % AM

Date:

Start

Stop

Dwell time s

Time:

Specimen:

Ambient temperature

°C

°C

f

Relative humidity

%

%

D min [unit]

Barometric pressure

kPa

kPa

D max [unit]

Frequency cycle

Cycle phase

Initial

During exposure

After

Load

Time

Start

Stop

Quantity [unit]

reference

indicated

Error [ v min ]

relative error [%] E

MPE [%]

Pass

Fail

Observed faults during exposure

Fault limit [%]

…………….

Frequency

Fault/Deviation

Significant

Acts on fault

MHz

Yes

No

Yes

No

Observations

Result

Pass

Fail


6.19  Span stability

6.19.1  Form 6.19.1 (3 runs) Span stability — measurement data for classes C and D

R 60-1, 5.7.2.6
R 60-2, 2.10.7.11
R 60-3, 2.2.9

Application no.:

Load cell model:

Serial no.:

E max :

n LC :

v min :

p LC : DR:

Force-generating system:

Indicating instrument:

Evaluator:

Conversion factor, f :

Minimum test load, D min :

Maximum test load, D max :

Note 

  1. Span is the result of subtracting the average indication at minimum test load from the average indication at maximum test load.

  2. Absolute (not relative) time shall be recorded.


Measurement no. 1:

Test load (unit)
( g , kg , t )
Run no. 1Run no. 2Run no. 3Average indication
(   )
Indication
(   )
TimeIndication
(   )
TimeIndication
(   )
Time
span

Evaluator: ____________________ Remarks:

Date:

Time:

Temperature: °C

Relative humidity: %

Barometric pressure: kPa


Measurement no. 2:

Test load (unit)
( g , kg , t )
Run no. 1Run no. 2Run no. 3Average indication
(   )
Indication
(   )
TimeIndication
(   )
TimeIndication
(   )
Time
span

Evaluator: ____________________ Remarks:

Date:

Time:

Temperature: °C

Relative humidity: %

Barometric pressure: kPa


Measurement no. 3:

Test load (unit)
( g , kg , t )
Run no. 1Run no. 2Run no. 3Average indication
(   )
Indication
(   )
TimeIndication
(   )
TimeIndication
(   )
Time
span

Evaluator: ____________________ Remarks:

Date:

Time:

Temperature: °C

Relative humidity: %

Barometric pressure: kPa


Measurement no. 4:

Test load (unit)
( g , kg , t )
Run no. 1Run no. 2Run no. 3Average indication
(   )
Indication
(   )
TimeIndication
(   )
TimeIndication
(   )
Time
span

Evaluator: ____________________ Remarks:

Date:

Time:

Temperature: °C

Relative humidity: %

Barometric pressure: kPa


Measurement no. 5:

Test load (unit)
( g , kg , t )
Run no. 1Run no. 2Run no. 3Average indication
(   )
Indication
(   )
TimeIndication
(   )
TimeIndication
(   )
Time
span

Evaluator: ____________________ Remarks:

Date:

Time:

Temperature: °C

Relative humidity: %

Barometric pressure: kPa


Measurement no. 6:

Test load (unit)
( g , kg , t )
Run no. 1Run no. 2Run no. 3Average indication
(   )
Indication
(   )
TimeIndication
(   )
TimeIndication
(   )
Time
span

Evaluator: ____________________ Remarks:

Date:

Time:

Temperature: °C

Relative humidity: %

Barometric pressure: kPa


Measurement no. 7:

Test load (unit)
( g , kg , t )
Run no. 1Run no. 2Run no. 3Average indication
(   )
Indication
(   )
TimeIndication
(   )
TimeIndication
(   )
Time
span

Evaluator: ____________________ Remarks:

Date:

Time:

Temperature: °C

Relative humidity: %

Barometric pressure: kPa


Measurement no. 8:

Test load (unit)
( g , kg , t )
Run no. 1Run no. 2Run no. 3Average indication
(   )
Indication
(   )
TimeIndication
(   )
TimeIndication
(   )
Time
span

Evaluator: ____________________ Remarks:

Date:

Time:

Temperature: °C

Relative humidity: %

Barometric pressure: kPa

6.19.2  Form 6.19.2 (5 runs) Span stability measurement data for class B

R 60-1, 5.7.2.6
R 60-2, 2.10.7.11
R 60-3, 2.2.9

Note 

  1. Span is the result of subtracting the average indication at minimum test load from the average indication at maximum test load.

  2. Absolute (not relative) time shall be recorded.

Table 25 — Measurement no. 1 (5 runs)
Test load (unit)
( g , kg , t )
Run no. 1Run no. 2Run no. 3Run no. 4Run no. 5Average indication
(   )
indication
(   )
Timeindication
(   )
Timeindication
(   )
Timeindication
(   )
Timeindication
(   )
Time
span

Evaluator: ____________________ Remarks:

Date:

Time:

Temperature: °C

Relative humidity: %

Barometric pressure: kPa


Table 26 — Measurement no. 2 (5 runs)
Test load (unit)
( g , kg , t )
Run no. 1Run no. 2Run no. 3Run no. 4Run no. 5Average indication
(   )
indication
(   )
Timeindication
(   )
Timeindication
(   )
Timeindication
(   )
Timeindication
(   )
Time
span

Evaluator: ____________________ Remarks:

Date:

Time:

Temperature: °C

Relative humidity: %

Barometric pressure: kPa


Table 27 — Measurement no. 3 (5 runs)
Test load (unit)
( g , kg , t )
Run no. 1Run no. 2Run no. 3Run no. 4Run no. 5Average indication
(   )
indication
(   )
Timeindication
(   )
Timeindication
(   )
Timeindication
(   )
Timeindication
(   )
Time
span

Evaluator: ____________________ Remarks:

Date:

Time:

Temperature: °C

Relative humidity: %

Barometric pressure: kPa


Table 28 — Measurement no. 4 (5 runs)
Test load (unit)
( g , kg , t )
Run no. 1Run no. 2Run no. 3Run no. 4Run no. 5Average indication
(   )
indication
(   )
Timeindication
(   )
Timeindication
(   )
Timeindication
(   )
Timeindication
(   )
Time
span

Evaluator: ____________________ Remarks:

Date:

Time:

Temperature: °C

Relative humidity: %

Barometric pressure: kPa


Table 29 — Measurement no. 5 (5 runs)
Test load (unit)
( g , kg , t )
Run no. 1Run no. 2Run no. 3Run no. 4Run no. 5Average indication
(   )
indication
(   )
Timeindication
(   )
Timeindication
(   )
Timeindication
(   )
Timeindication
(   )
Time
span

Evaluator: ____________________ Remarks:

Date:

Time:

Temperature: °C

Relative humidity: %

Barometric pressure: kPa

Table 30 — Measurement no. 6 (5 runs)
Test load (unit)
( g , kg , t )
Run no. 1Run no. 2Run no. 3Run no. 4Run no. 5Average indication
(   )
indication
(   )
Timeindication
(   )
Timeindication
(   )
Timeindication
(   )
Timeindication
(   )
Time
span

Evaluator: ____________________ Remarks:

Date:

Time:

Temperature: °C

Relative humidity: %

Barometric pressure: kPa


Table 31 — Measurement no. 7 (5 runs)
Test load (unit)
( g , kg , t )
Run no. 1Run no. 2Run no. 3Run no. 4Run no. 5Average indication
(   )
indication
(   )
Timeindication
(   )
Timeindication
(   )
Timeindication
(   )
Timeindication
(   )
Time
span

Evaluator: ____________________ Remarks:

Date:

Time:

Temperature: °C

Relative humidity: %

Barometric pressure: kPa


Table 32 — Measurement no. 8 (5 runs)
Test load (unit)
( g , kg , t )
Run no. 1Run no. 2Run no. 3Run no. 4Run no. 5Average indication
(   )
indication
(   )
Timeindication
(   )
Timeindication
(   )
Timeindication
(   )
Timeindication
(   )
Time
span

Evaluator: ____________________ Remarks:

Date:

Time:

Temperature: °C

Relative humidity: %

Barometric pressure: kPa

6.19.3  Form 6.19.3 Span stability — summary of test results

R 60-1, 5.7.2.6
R 60-2, 2.10.7.11
R 60-3, 2.2.9

Application no.:

Load cell model:

Serial no.:

E max :

n LC :

v min :

p LC : DR:

Force-generating system:

Indicating instrument:

Evaluator:

Table 33 — Table 6.19.3
Measurement no.SpanVariation
( v min )
Maximum allowable variation ( v min )
(     )( v min )
1
2
3
4
5
6
7
8

PASS: ☐ FAIL: ☐

Remarks: