2.10.7.9 Exposure to radiated RF electromagnetic fields (see R 60-1, 5.7.2.5 Disturbances)

This test is conducted to verify compliance with the provisions in R 60-1, 5.7.2.5 under conditions of exposure to electromagnetic fields.

Applicable standard:

IEC 61000-4-3 [22]: No. 6 (test generator), No. 7 (test set-up), No. 8 (test procedure)

Test method

Exposure to specified electromagnetic fields

Test conditions

In accordance with 2.7.3.1 Environmental conditions

Preconditioning of the load cell

Stabilise the load cell under constant environmental conditions.

Test procedure in brief

The load cell is exposed to electromagnetic fields with the required field strength and the field uniformity as defined in the referred standard.

The level of field strength specified refers to the field generated by the unmodulated carrier wave.

The load cell shall be exposed to the modulated wave field. The frequency sweep shall be made only pausing to adjust the RF signal level or to switch RF-generators, amplifiers and antennas if necessary. Where the frequency range is swept incrementally, the step size shall not exceed 1 % of the preceding frequency value.

The dwell time of the amplitude modulated carrier at each frequency shall not be less than the time necessary for the load cell to be exercised and to respond, but shall in no case be less than 0.5 s.

Adequate EM fields can be generated in facilities of different type and set-up the use of which is limited by the dimensions of the load cell and the frequency range of the facility.

Test load

The test shall be performed with no load or with one small test load ().

Test levels

Level 3:

Frequency range: a)

to ;

Field strength: ;

Modulation: 80 % AM, sine wave.