2.10.6.11 Conduct damp heat, steady state test
This test is conducted to verify compliance with the provisions in R 60-1, 5.6.1 or R 60-1, 5.6.3 under conditions of high humidity and constant temperature.
Applicable standards:
IEC 60068-2-78: Environmental testing — Part 2: Tests. Test Ca: Damp heat, steady state,
Environmental testing — Part 2: Tests. Test Cb: Damp heat, steady state, primarily for equipment [12].
IEC 60068-3-4: Environmental testing — Part 2: Tests. Guidance for damp heat tests [11].
Test method | Exposure to damp heat in steady state |
Test conditions | Relative humidity of 85 % |
Preconditioning of load cell | Place the load cell in the chamber with the output connection external to the chamber, and switched on. |
Test procedure in brief | This test involves exposure of the load cell to a constant temperature and a constant relative humidity. The load cell shall be tested as specified in 2.10.6.1 to 2.10.6.10:
|