2.10.5.12 Conduct damp heat, cyclic test (CH)
This test is conducted to verify compliance with the provisions in R 60-1, 5.6.3.1 under conditions of high humidity combined with cyclic temperature changes
Applicable standards:
IEC 60068-2-30 [10]: Environmental testing – Part 2: Tests
Test Db and guidance: Damp heat cyclic (12 + 12-hour) cycle
IEC 60068-3-4 [11]: Environmental testing — Part 2: Tests. Guidance for damp heat tests
Test method | Exposure to damp heat with cyclic temperature variation |
Test conditions | The relative humidity is between 80 % and 96 % and the temperature is varied from to , in accordance with the specified cycle. |
Preconditioning of load cell | Load cell placed in the chamber with the output connection external to the chamber, and switched off. Use variant 2 of IEC 60068-2-30 Ed. 3.0 (2005-08) when lowering the temperature. |
Initial measurements | |
Test procedure in brief | This test consists of exposure to 12 temperature cycles of 24-hour duration each. Condensation is expected to occur on the load cell during the temperature rise. The 24 h cycle comprises:
The stabilising period before and recovery period after the cyclic exposure shall be such that the temperature of all parts of the load cell is within of its final value. Recovery conditions and final measurements: According to 2.10.5.13 to 2.10.5.15 . |