1.2 Measurement standards
The expanded uncertainty, (for coverage factor ), for the combination of the force-generating system and the indicating instrument used during the tests to observe the load cell output shall be less than 1/3 times the MPE of the load cell under test [OIML G 1-100 Guide to the Expression of Uncertainty in Measurement] [7].