5.6.3.1 Humidity error — CH or unmarked load cells
This requirement is only applicable to load cells marked CH or with no humidity symbol marking and not applicable to load cells marked NH or SH.
The influence of exposure to temperature cycles specified in R 60-2, 2.10.5.12 R060-2 / current-metanorma-collection/R060-2, 2.10.5.12 on the load cell output for minimum load shall not be greater than 4 % of the difference between the output on the maximum capacity, , and that at the minimum dead load, .
The influence of exposure to temperature cycles specified in R 60-2, 2.10.5.12 on the load cell output for the maximum load shall not be greater than the load cell verification interval, v.