5.3 Temperature measurement

Appropriate temperature-measuring devices shall be used to ensure accurate temperature correction of the liquid volume. These devices shall be mounted in suitable locations on the prover and the measuring system. It is recommended to use the temperature-measuring devices with an expanded uncertainty of or less. Calibration certificates shall be provided for the devices.